Abstract

This work reports that the intensity ratio of spectral lines having similar self-absorption characteristics during laser induced breakdown spectroscopy (LIBS) analysis can become nearly constant over a wide range of irradiation conditions if the intensities are integrated over a sufficiently long time. It is shown that the plasma temperature and intensity ratio of these spectral lines have temporal similarity. The spectral lines with similar self-absorption properties may be selected to improve the accuracy and consistency of LIBS analysis results under an environment with fluctuating measurement conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.