Recent revision of exposure and air kerma standards in Standards Laboratories require a simultaneous change in physical parameters at other positions in the dosimetry chain. Adoption of new data, recommended by international organizations, will introduce changes in absorbed dose determinations in high-energy photon beams using ionization chambers. A new code of practice has therefore been drafted using a consistent set of data. In this code of practice, single conversion factors are given to convert ionization chamber reading to absorbed dose to water for some types of reference ionization chamber as a function of radiation quality. Equations and recommended numerical data for the physical parameters and correction and conversion factors will be provided.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.