Abstract

The unique considerations to determine the number of particles per liter and analyte mass (particle size) distributions from single particle ICP-MS measurements with quadrupole or sequential sector field mass spectrometers are discussed. The short, transient nature of signals from single particles affects the optimum selection of dwell times and pulse counting versus analog detection. The signal peak width due to a cloud of ions produced from a single particle also limits the maximum number of single particles that can be accurately measured in a given total measurement time and the dynamic range of the particle number concentration. The range of analyte mass (particle size) measurements is limited by the rates of vaporization and diffusion. Approaches to calibrate the number of particles detected as a function of the number of particles per mL and to calibrate the signal from a single particle as a function of analyte mass are discussed. Measurement of particle diameter distributions for spherical, solid particles are assessed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.