Abstract

Acoustic pressure can be measured with a hydrophone. Hydrophone measurements can underestimate incident acoustic pressure due to spatial averaging effects across the hydrophone sensitive element. The spatial averaging filter for a nonlinear focused beam is a low-pass filter that decreases monotonically from 1 to 0 as frequency increases from 0 to infinity. Experiments were performed to test an analytic model for the spatial averaging filter. Nonlinear pressure tone bursts were generated by three source transducers with driving frequencies ranging from 2.5 to 6 MHz, diameters ranging from 19 to 64 mm, and focal lengths ranging from 38 to 89 mm. The nonlinear pressure fields were measured using four needle hydrophones with nominal geometrical sensitive element diameters of 200, 400, 600, and [Formula: see text]. The average root-mean-square difference between theoretical and experimental spatial averaging filters was 5.8% ± 2.6%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.