Abstract

This paper presents the characterisation performed at IRSN (France) of an H(p)(10) chamber in terms of calibration coefficient and correction factors for the radiation qualities of ISO narrow spectrum series. The chamber response, expressed in H(p)(10) using conversion coefficients h(p)(K)(10; N, alpha) listed in ISO 4037-3 in the energy range from 30 to 1250 keV and for angles of incidence between 0 and 70 degrees, was found to be within approximately 10%. However, for photon energies <30 keV, an overresponse of the chamber that could reach 100% was observed. Nevertheless, this overresponse was reduced to 25% using the conversion coefficients estimated at Physikalisch-Technische Bundesanstalt (PTB). This implies that the X-ray spectra produced by the IRSN X-ray units are very similar to those produced by PTB, both containing a little bit more high-energy photons than the spectra used in ISO 4037-3. The dose rate dependence of the chamber tested by gamma radiation from (60)Co sources was found to be within 2% in the range of 0.3 mSv h(-1) to 1 Sv h(-1). The H(p)(10) chamber can measure directly the conventional true value of H(p)(10) after calibration by a reference laboratory, and can be used for transferring H(p)(10) reference quantities from a reference laboratory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.