Abstract

In high-energy photon reference fields the value of the air kerma rate is determined by using ionization chambers (ICs). From the charge collected inside the IC the dose can be calculated using a set of calibration and correction factors according to ISO 4037-2. A crucial parameter is the correction for the attenuation and scattering of the primary radiation due to the chamber wall. This parameter can be determined using Monte Carlo calculations. The evaluation of the factor was performed for a commercially available IC of the type Victoreen 550-3 under different build-up conditions. The results were verified by measurements in the R-F high-energy photon fields according to ISO 4037-1 at the Physikalisch-Technische Bundesanstalt (PTB) and the Japan Atomic Energy Agency (JAEA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.