Abstract

Pulse tube (PT) cryocoolers can be used to cool SQUIDs, however their electromagnetic interference (EMI) through the conducted high frequency noise currents are normally too large to operate SQUIDs. In this paper, high frequency noise currents of a pulse tube cryocooler system are simultaneously observed at several critical points to analyze conducting paths of the noise currents and to find a solution for this problem. High frequency noise currents are detected by current sensors consisting of a current transformer and a terminating resistor at the secondary side. By covering three-phase power lines and a neutral line connecting the valve motor to the inverter with a copper mesh, and by using it as an additional neutral line, conducted EMI (noise currents) to the cryostat has bee reduced by one fifth. With this treatment, SQUIDs have been operated normally.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.