Abstract

A high-performance liquid chromatographic method for the simultaneous determination of thiamin and thiamin phosphate esters in human blood or serum has been developed. The eluent consists of acetonitrile and phosphate buffer, in the ratios 90:10 (v/v) for the elution of thiamine and 60:40 (v/v) for the phosphate esters. The four compounds are eluted within 15 min. The detection limit is 13-16 fmol. Between-assay variation is 5-11%. Samples of whole blood and serum from 30 healthy adults were analysed. The following reference values were obtained for 15 females 15 males (nM, mean +/- S.D.). In serum: thiamin, 10.9 +/- 2.9/16.9 +/- 3.3; thiamin monophosphate, 8.3 +/- 1.5/3.7 +/- 1.5. In whole blood: thiamin, 29.6 +/- 10.0/33.4 +/- 10.4; thiamin monophosphate, 9.7 +/- 2.3/10.9 +/- 5.1; thiamin diphosphate, 121 +/- 29.6/165 +/- 40.4.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.