Abstract

Two publications on interferometric gauge block measurements claiming extraordinarily low uncertainties gave rise to some critical considerations. For such interferometric measurements, wringing of the gauge block to a platen is a basic element in the measurement process. Among the effects limiting the attainable uncertainty level, the elastic deformation of the gauge block by the wringing process has to be included in the evaluation model and to be appropriately taken into account in the uncertainty budget.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.