Abstract

A new differential pumping design for external source Fourier transform ion cyclotron resonance mass spectrometry is described. A network of concentric tubes of increasing diameter terminates at a series of conductance limits across which a pressure from atmosphere to low-10 −8 torr is achieved. This design permits high-pressure sources to be positioned within the solenoidal superconducting magnet less than 20 cm from the analyzer trapped ion cell. Ionization at high magnetic field offers the advantage of radial ion confinement and consequently delivers enhanced ion current to the trapped ion cell. Ion injection utilizing this vacuum chamber design is simpler than previously reported serial pumping stage designs because elaborate focusing optics to overcome the magnetic mirror effect are unnecessary. Two probe-mounted atmospheric pressure sources are described as evidence of the general applicability of the concentric tube vacuum chamber. An electrospray source that delivers several hundred picoamperes of ion current to the cell yields high-sensitivity spectra of proteins beyond 100 kDa. Improved pumping compared with a prototype concentric tube network configuration now permits mass resolution in excess of 20,000 for the [M + 4H] 4+ ion of melittin. The resolution is sufficient to distinguish isotope peaks within a single charge state. A probe-mounted, pulsed-laser ablation source that permits cluster formation in the strong magnetic field is also demonstrated.