Abstract
Evaluate the agreement between computerized QT/QTc measurements from the bedside monitor (four leads) and a time-matched standard 12-lead ECG. Prospective observational study in three adult ICUs. QT/QTc measurements were obtained from a convenience sample, and the two ECG types were ≤ 30 min apart. Agreement was evaluated using Bland-Altman analysis. A total of 120 patients were evaluated for inclusion, and 60 (50%) had a 12-lead ECG for comparison. The mean bias difference for QT measurements was not statistically different (β = -2.47, 95% CI = 5.50 to -11.05; p = 0.44; limits of agreement (LOA) = -64.37 to 59.44). Similar non-statistical differences were observed for QTc (β = -3.20, 95% CI = 5.50 to -11.05; p = 0.44; LOA = -67.43 to 61.03). There was good agreement for both QT and QTc measurements between the two methods. These pilot data are promising and suggest QT/QTc measurements from bedside monitors (four leads) may be an acceptable alternative to obtaining additional standard 12-lead ECGs. Given that half of the ICU patients screened did not have a 12-lead ECG recorded, bedside monitor QT/QTc's could identify at-risk patients. However, an evaluation in a larger sample and non-ICU patients is warranted.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.