Abstract

The purpose of this study is to assess QT interval measurements from the EASI 12-lead ECG as compared to the standard 12-lead ECG. The QT interval was automatically measured in simultaneously recorded standard and EASI 12-lead ECGs, using a validated wavelet-based delineator. The agreement between the two sets of measurements was quantified both on a lead-by-lead basis and a multilead basis. The results show an acceptable agreement between QT measurements in the two lead systems, with correlation coefficients (CC) 0.91-0.98 depending on the lead. Standard deviations range from 9.2 ms to 26.4 ms depending on the selected lead. In a few patients large inter-system differences were found, mainly due to different T wave morphologies. Using a multilead delineation, QT intervals were considerably more stable than single-lead measurements and resulted in a much better agreement between the two lead systems (CC: 0.98, QT difference: 1.1 ms +/- 9.8 ms). Thus, EASI-derived 12-lead ECG may be used for reliable QT monitoring when the multilead delineation approach is adopted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.