Abstract

Autologous arteriovenous fistula (AVF) is a commonly used vascular access (VA) for hemodialysis, and hemodynamic changes are one of the main factors for its failure. To explore the effect of geometry on the hemodynamics in the AVF, a modified model is built with a gradual and smooth turn at the anastomosis and is compared with the traditional model, which has an abrupt sharp turn at the anastomisis. Transient computational fluid dynamics (CFD) simulations were performed for the comparison and analysis of the hemodynamic fields of the two models at different stages of the pulse cycle. The results showed that the low shear stress region and high oscillatory shear stress region in the modified AVF model coincided with regions of intimal hyperplasia that have been identified by previous studies. A comparison with the blood flow velocities measured in vivo was performed, and the error between the simulation results and the medical data was reduced by 22% in the modified model, which verifies the rationality and utility of the modified model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.