Abstract
Using a multilevel fast multipole method, coupled with the shadow imaging of femtosecond photoelectron pulses for validation, we quantitatively elucidate the photocathode, space charge, and virtual cathode physics, which fundamentally limit the spatiotemporal and spectroscopic resolution and throughput of ultrafast electron microscope (UEM) systems. We present a simple microscopic description to capture the nonlinear beam dynamics based on a two-fluid picture and elucidate an unexpected dominant role of image potential pinning in accelerating the emittance growth process. These calculations set theoretical limits on the performance of UEM systems and provide useful guides for photocathode design for high-brightness electron beam systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.