Abstract

The Comprehensive Nuclear-Test-Ban Treaty (CTBT) is supported by a network of certified laboratories that perform high-resolution gamma-spectrometry on global air filter samples for the identification of 85 radionuclides. At the UK CTBT Radionuclide Laboratory (GBL15), the use of advanced Compton suppressed systems has been investigated to reduce the Compton continuum and improve detection sensitivity. Samples collected from the Philippines and during the Fukushima incident have been measured, demonstrating Compton continuum reductions of 28–59 % with suppression factors of 0.1–147.0. Detection sensitivity has been improved with typically 40 % lower MDAs, including 140Ba to meet CTBT requirements. True coincidence summing effects have been considered, including the application to remove interferences by the elimination of gamma-rays in cascade. This has been demonstrated for the removal of 134Cs allowing improved 131I measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.