Abstract

Optical coherence tomography (OCT) is capable of angstrom-scale vibrometry of particular interest to researchers of auditory mechanics. We develop a method for compressed sensing vibrometry using OCT that significantly reduces acquisition time for dense motion maps. Our method, based on total generalized variation with uniform subsampling, can reduce the number of samples needed to measure motion maps by a factor of ten with less than 5% normalized mean square error when tested on a diverse set of in vivo measurements from the gerbil cochlea. This opens up the possibility for more complex in vivo experiments for cochlear mechanics.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call