Abstract

The electrochemical behaviour of the Ti–13Nb–13Zr and Ti–6Al–4V ELI alloys with martensitic microstructures was investigated by polarization and electrochemical impedance spectroscopy (EIS) in Ringer’s solution. The impedance spectra were interpreted by a two time-constants equivalent circuit. Both investigated alloys showed high corrosion resistance, but the thin and uniform passive film on the Ti–6Al–4V ELI alloy surface was more protective. The inner barrier and outer porous layer were highly resistant and capacitive. However, thicker and more porous passive film on the Ti–13Nb–13Zr alloy surface may be beneficial for osteointegration. The suitable thermomechanical processing improved the corrosion resistance of Ti–13Nb–13Zr alloy.