Abstract

Purpose – The purpose of this paper is to investigate the influence of composite coating structure on the reliability of adhesive flip chip joints. The need for conformal coating is considered, especially for medical applications, and medical sterilization is also considered.Design/methodology/approach – Two test lots were assembled and one of them was sterilized using gamma sterilization. Both test lots were coated first with epoxy and then with Parylene C, resulting in a composite coating structure. The reliability was studied using a constant humidity test and the failure analysis was performed with cross‐sections and scanning electron microscopy analysis. These results were compared to earlier research results on conformal coatings.Findings – The reliability of both test lots proved to be good. The composite coating structure shields the joints from humidity and improves the reliability compared to non‐coated test samples. When the conformal coating was compared to the pure Parylene C coated test lot,...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.