Abstract
The aim of this study was to analyse the short- and medium-term complications of the Motec total wrist arthroplasty (TWA). Identifying exact modes of failure and their causes should allow surgeons to avoid or mitigate these risks in the future. Retrospective analysis of prospectively collected data from six hand surgeons at five international centres provided details of 171 Motec TWAs. The mean follow-up was 5.8 years (range 18 months to 12 years). There were 33 (19%) complications within our cohort, with a revision rate of 8.2% (14 revisions). There was no difference in complication rates between metal-on-metal and metal-on-polymer articulations. Failure of osseointegration was the most common complication. Problems with soft tissue balancing, implant impingement related osteolysis, bony impingement and metacarpal fracture were found to be other preventable causes of failure in this series. Elimination of these preventable complications will improve survival rates for this implant.Level of evidence: IV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.