Abstract

The selectivity of voltammetric measurements following extractive and adsorptive accumulation at carbon paste electrode is evaluated. Various situations where more than one electroactive species accumulate at the electrode are considered. It is shown that extractable species can be measured in the presence of adsorbable species if the outer layer of the surface is removed following the accumulation step and the voltammetric measurement is performed in a blank solution. As the extractive accumulation process can be treated on the basis of solvent extraction principles, the voltammetric measurement, performed on the electrode interior, provides the desired quantitative information. Due to differences in acid-base properties of various extractable organic compounds, it is possible to design an accumulation scheme specific for a given analyte. New information on the nature of the extractive accumulation step is reported. Chlorpromazine, uric acid, N,N -dimethylaniline and butylated hydroxyanisole are used as test systems. The implications of the method on the development of selective voltammetric sensors are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.