Abstract

We are developing a depth-encoding Anger detector for imaging 511 keV annihilation photons. Our design integrates wavelength-shifting (WLS) fibers onto an Anger-type detector to allow measurement of depth-of-interaction (DOI). As few as 0.3% of blue scintillation photons created in NaI(Tl) contribute to detectable signal at the end of the WLS fiber ribbon. Thus, maximizing signal output from the fibers is important for achieving the best possible DOI measurement. We have investigated the effects of two design options on the signal output from the WLS fibers: fiber geometry and methods of assembling fibers into a ribbon. A blue-to-green WLS fiber (Saint Gobain BCF-91A) was chosen because its absorption spectrum matches well the emission spectrum of NaI(Tl). Its emissions, peaked at 500 nm, can be detected with reasonable quantum efficiency by enhanced-green-response PMTs. These fibers are available in several configurations (round or square cross-section, single or multiple cladding layers) and sizes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.