Abstract

This investigation focused on direct comparison of two popular multidimensional liquid-gas chromatography (LC-GC) systems, the Y-interface (retention gap approach) and the syringe-based interface (programmed temperature vaporizer approach). Such transfer devices are structurally very different, and could potentially have a substantial effect on the outcome of a specific application. In this work the application was a topic of much current interest, determination of mineral oil saturated hydrocarbon (MOSH) contamination of a series of food products (rice, pasta, icing sugar, olive oil); the final results were then compared. The two LC-GC methods developed were validated for linearity over the calibration range, analyte discrimination, precision, accuracy, and limits of detection and quantification. No significant differences were found between the two approaches.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.