Abstract

Univariate and multivariate (partial least squares 1, PLS1) calibration techniques are compared for the determination of Pt, Pd and Rh in a complex sample matrix. The univariate techniques utilised either pure standards, pure standards with interelement correction factors applied, or matrix matched standards. Univariate calibration yielded relative root mean square errors (RRMSEs) of prediction for Pt, Pd and Rh of 19, 15, and 54% and 13, 18, and 88% for test solutions and real samples, respectively. Univariate calibration with matrix matching proved the most accurate method with RRMSEs for Pt, Pd and Rh of 2.5, 3.7 and 2.4% for a series of synthetic test solutions, and 12, 2.3 and 8.0% for real samples, respectively. In comparison, the multivariate calibration method yielded relative root mean errors for Pt, Pd and Rh of 5.8, 3.0 and 3.5% in the test solutions, and 32, 7.4 and 76% in the real samples. The relative error for the matrix matching and PLS1 techniques was dependent upon analyte concentration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.