Abstract

In order to compare the humidity standards at the Centre Technique des Industries Aerauliques et Thermiques (France) and the Centre for Metrology and Accreditation (Finland), a dew-point hygrometer was calibrated at both laboratories within a EUROMET project. There were eight measurement points in the range -40 °C to 60 °C. The difference between the standards was determined and its uncertainty was evaluated. The difference was between -0,03 °C and +0,09 °C with the average being 0,01 °C. The uncertainty of the difference was ±0,08 °C to ±0,17 °C (k = 2). The measurements and the analysis of the results are reported in this paper.