Abstract

Abstract At Centre for Metrology and Accreditation (MIKES), a primary calibration system for gas mass flows between 0.42 mg/s and 625 mg/s is based on dynamic weighing providing traceability directly to the national mass and time standards. To evaluate the agreement of the system with a volumetric primary calibrator, the international comparison between MIKES and the Swiss Federal Office of Metrology and Accreditation (METAS, Switzerland) was carried out. At METAS, the primary low gas flow standard provides traceability to the Swiss national measurement standards for length and time. The results of the comparison showed that the two systems do not deviate more than ±0.15% in the mass flow range 0.42–625 mg/s.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.