Abstract

Cobalt (Co) alloys are used extensively in a wide range of medical devices due to their biocompatibility, durability, and properties. In 2017 the European Chemical Agency (ECHA) proposed to classify Co metal as a category 1B carcinogen (i.e. presumed to have carcinogenic potential for humans), as a reproductive hazard category 1B (i.e. presumed human reproductive toxicant) and as a category 2 mutagen. Even more, the European Medical Device Regulation (MDR) requires that medical devices contain > 0,1 w/w of substances that are category 1 A and 1B (CMR). As far as the European Medical Device Regulation, it is not specified if the Co content and concentration had to be determined/measured on the entire product or only on the product components. The object of this work is the comparison of Co profile in three different suppliers of stainless-steel needles as is and after being processed (i.e. sterilized) and then provide the proper interpretation and application of MDR requirements. The study and the extractable profile demonstrate the low cobalt content on needles and on a total syringe, thus suggesting to consider the Co content only on the syringe component.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.