Abstract
Sap nitrate was determined with the ion-selective electrodes, HACH combination nitrate electrode, and the CARDY nitrate meter on pepper petioles over five sample dates (n = 160). The electrode values were compared to the Cd reduction method performed on a Lachat automated ion analyzer with flow injection analysis. Thirty petioles were collected from plots of each of several N rate studies and the sap expressed by a hand-held garlic press. Correlation among the techniques were similar (r > 0.9), but the CARDY meter constantly read 100 to 175 ppm higher than the HACH. Across all dates the standard deviation of the difference, compared with the Cd reduction, for the HACH = 16 ppm, while for the CARDY it was 50. While the CARDY meter is easier to use and has fewer steps, the HACH electrode values were closer to the true readings and less erratic. One must use care when interpreting nitrate sufficiency value ranges with different quick-test techniques.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.