Abstract

Efficient phosphoproteomic analysis of small amounts of biological samples (e.g. tissue biopsies) requires carefully selected enrichment and purification steps prior to the nanoflow HPLC-MS/MS analysis. Solid-phase extraction (SPE) is one of the most commonly used approaches for sample preparation. Several stationary phases are available for peptide SPE purification, however, most of the published methods are not optimized to provide good recoveries of phosphorylated peptides. Our goal was to investigate the performance of 13 self-packed and 3 commercial centrifugal SPE cartridges/spin tips, thus enhancing the efficiency of the phosphoproteomic analysis of small amounts of complex protein mixtures. Eight reversed-phase (RP), five graphite, two ion-exchange, and one hydrophilic-lipophilic balance (HLB) stationary phase were evaluated. Two RP, one graphite, and the HLB self-packed centrifugal SPE tips provided excellent results for the purification of 1 µg tissue and cell line digests. Using these methods, the sample loss was significantly reduced compared to one of the commercial SPE methods, 22-58% more unique phosphopeptides were identified, and the recovery was higher by 132-155%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.