Abstract

We evaluated a commercial enzyme immunoassay (EIA) kit for detection of Legionella pneumophila serogroup 1 soluble antigen by comparing it to radioimmunoassay (RIA), using both concentrated and nonconcentrated urine samples. The sensitivity of EIA was 67.4% in nonconcentrated urine samples and 82.6% in concentrated urine samples. The sensitivity of RIA was 60.9% and 84.8% in nonconcentrated and concentrated urine samples, respectively. Our study indicates that the sensitivity and specificity of EIA are comparable to those of RIA, and that concentrating the antigen by selective ultrafiltration increases sensitivity for both EIA and RIA, with no significant decrease in specificity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.