Abstract
Given that the rate of mortality by Legionella pneumonia increases in incorrectly treated patients, rapid diagnosis and early antibiotic treatment are needed. We have assessed the performance of a new enzyme immunoassay (EIA) test (Bartels Inc. Trinity Biotech Company, Wicklow, Ireland) to detect Legionella pneumophila antigen in urine comparing it to Binax EIA (Binax, Portland, Maine). We also evaluated the capability of both EIAs to detect extracted soluble antigens of Legionella strains. Using nonconcentrated urine samples (NCU) the sensitivity of Bartels EIA was 74.1% (66/89) and the sensitivity of Binax EIA was 51.7% (46/89). The sensitivity of both EIA tests were 91.5% (54/59) using concentrated urine samples (CU). Specificity of both EIA tests was 100% in NCU and CU. Bartels EIA was able to detect all serogroup L. pneumophila antigens, achieving a higher sensitivity in the case of L. pneumophila serogroup 1 soluble antigen. The new EIA was found to be a useful test for the rapid diagnosis of Legionella pneumonia, being a better alternative to the Binax EIA if NCU is used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.