Abstract
The aim of this study was to compare osseointegration and surface characteristics of zirconia implants made by the powder injection molding (PIM) technique against those made by the conventional milling procedure in rabbit tibiae. Surface characteristics of 2 types of implants were evaluated. Sixteen rabbits received 2 types of external hex implants with similar geometry, either machined zirconia implants or PIM zirconia implants, in the tibiae. Removal torque tests and histomorphometric analyses were performed. The roughness of the PIM zirconia implants was higher than that of machined zirconia implants. The PIM zirconia implants exhibited significantly higher bone-implant contact and removal torque values than the machined zirconia implants (p<0.001). The osseointegration of the PIM zirconia implant is promising, and PIM, using the roughened mold etching technique, can produce substantially rougher surfaces on zirconia implants.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.