Abstract
Central systolic blood pressure (cSBP) can be derived by the general transfer function of the radial pressure wave, as used in the SphygmoCor device, or by regression equation from directly measured late systolic shoulder of the radial pressure wave (pSBP2), as used in the Omron HEM-9000AI device. The aim of this study was to compare the SphygmoCor estimates of cSBP with 2 estimates of cSBP provided by the Omron HEM-9000AI (cSBP, pSBP2) in a large cohort of the white population. In 391 patients aged 52.3±13.5 years (46% men) from the Czech post-MONICA Study, cSBP was measured using the SphygmoCor and Omron HEM-9000AI devices in random order. Omron cSBP and pSBP2 were perfectly correlated (r = 1.0; P < 0.0001). There was a strong correlation (r = 0.97; P < 0.0001) between Omron and SphygmoCor cSBP estimates, but Omron estimate was 13.1±4.7mm Hg higher than SphygmoCor cSBP. On the other hand, Omron pSBP2 strongly correlated with SphygmoCor cSBP (r = 0.97; P < 0.0001) and was 1.7±4.2mm Hg lower than SphygmoCor cSBP. In multivariable analysis, anthropometric and cardiovascular risk factors explained only 10% of the variance of the cSBP difference between devices while explaining 52% of the systolic blood pressure amplification variance. Estimation of cSBP based on the late systolic shoulder of the radial wave provides a comparable accuracy with the validated general transfer function. When comparing Omron HEM-9000AI and SphygmoCor estimates of cSBP, Omron pSBP2 should be used. The difference between both devices in cSBP may be explained by differences in calibration.
Published Version (
Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have