Abstract

The Josephson array voltage standard at the Beijing Institute of Radio Metrology and Measurement (BIRMM) has been used in China to realize the volt since 1992. In order to verify the accuracy and reliability of the voltage standard at the BIRMM, a comparison of the Josephson array voltage standards at 1,018 V and 10 V between the BIRMM, the Danish Institute of Fundamental Metrology (DFM) and the National Physical Laboratory (NPL), UK, has been performed. Two Zener reference standards were used to transfer the voltage between the three laboratories. The results show a difference of less than 2 × 10-8.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.