Abstract

We have compared the Josephson array voltage standard (JAVS) held by the Singapore Institute of Standards and Industrial Research (SISIR) with that of the Danish Institute of Fundamental Metrology (DFM). The comparison took place in Singapore from 20-25 March 1995. A commercial Zener reference voltage standard was used as transfer instrument between the two laboratories. The result of the comparison was an observed difference of 13 nV with a combined uncertainty of 9 nV. The difference was independent of several measurement configurations used.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.