Abstract
This study aimed to examine the effect of the weight initializers on the respiratory signal prediction performance using the long short-term memory (LSTM) model. Respiratory signals collected with the CyberKnife Synchrony device during 304 breathing motion traces were used in this study. The effectiveness of four weight initializers (Glorot, He, Orthogonal, and Narrow-normal) on the prediction performance of the LSTM model was investigated. The prediction performance was evaluated by the normalized root mean square error (NRMSE) between the ground truth and predicted respiratory signal. Among the four initializers, the He initializer showed the best performance. The mean NRMSE with 385-ms ahead time using the He initializer was superior by 7.5%, 8.3%, and 11.3% as compared to that using the Glorot, Orthogonal, and Narrow-normal initializer, respectively. The confidence interval of NRMSE using Glorot, He, Orthogonal, and Narrow-normal initializer were [0.099, 0.175], [0.097, 0.147], [0.101, 0.176], and [0.107, 0.178], respectively. The experiment results in this study indicated that He could be a valuable initializer in the LSTM model for the respiratory signal prediction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.