Abstract

The common methods for patient dose estimations in computed tomography (CT) are thermoluminescence dosemeter (TLD) measurements or the usage of software packages based on Monte Carlo simulations like CT-Expo or the newer CTVoxDos, which uses the ICRP Reference Adult Male (ICRP 110). Organ (OD) and effective doses of a CT protocol of the upper abdomen are compared. Compared to CTVoxDos, ODs inferred by TLD measurement using an anthropomorphic phantom differ by $\mathbf{(19\pm 16)\,\%}$ inside the primary radiation field, $\mathbf{(14\pm 2)\,\%}$ for partially primary irradiated organs and $\mathbf{(34\pm 38)\,\%}$ in the scattered radiation field. ODs estimated by CT-Expo show a mean deviation of $\mathbf{(16\pm 9)\,\%}$ (primary irradiated) and $\mathbf{(28\pm 31)\,\%}$ (scatter irradiated) from ODs estimated by CTVoxDos.