Abstract
In recent years, silicon photomultipliers (SiPMs) became popular candidates for applications in positron emission tomography (PET) scanners or PET based multi-modality detectors. In this paper, we perform a comparative study of digital and analog SiPM in gamma spectroscopy with a lutetium-yttrium oxyorthosilicate (LYSO) scintillator. The energy spectrum in terms of photo-electron numbers (fired pixels) is obtained directly from the digital SiPM. The energy spectrum obtained with analog SiPM is converted to number of fired pixels using single photoelectron spectrum measurement. Non-linear behavior of both detectors is studied. Energy response for 511 keV gamma ray under non-linear conditions for both detectors are corrected using a Monte-Carlo simulation tool.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.