Abstract

This study addresses the comparative pull-out strengths of cortical (4.5 mm) and fully threaded cancellous (6.5 mm) bone screws from sites about the pelvis commonly used in the fixation of posterior column acetabular fractures and sacroiliac (SI) disruptions. These sites include one from lateral to medial through the posterior column, two from posterolateral to anteromedial across the SI joint, one vertically through the sacral ala, and one vertically through the iliopubic column. Statistical analysis showed no significant difference in the force required for pull-out failure of the cortical versus the cancellous screws at any of the sites tested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.