Abstract

To determine whether 2.0 mm cancellous screws are superior to 2.0 mm cortical screws when inserted into cancellous and bicortical bone. Biomechanical study. The 2.0 mm cancellous screws and 2.0 mm cortical screws were inserted according to the recommended guidelines in synthetic cancellous and bicortical blocks. Fifteen screw-block constructs per group were tested to failure in axial pullout. Axial pullout strength and yield strength were calculated. Data were analyzed using a one-way ANOVA. The 2.0 mm cortical screws achieved lower axial pullout strength than 2.0 mm cancellous screws in cancellous blocks. The 2.0 mm cortical screws achieved greater pullout strength than 2.0 mm cancellous screws in bicortical blocks. The 2.0 mm cancellous screws may offer a biomechanical advantage in bone with thin cortices (<1 mm thick), whereas 2.0 mm cortical screws may be preferred in cortical bone with cortices measuring at least 1 mm in thickness.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.