Abstract

ABSTRACTThis paper discusses metrologically the best practice regarding the calibration curves applied to H2S mass fractions determination in natural gas by gas chromatography with sulfur chemiluminescence detection (GC-SCD). Three calibration curves were constructed by performing GC-SCD analysis of different H2S gas standard concentrations (from 3 mg kg−1 up to 500 mg kg−1). These experimental curves are better fitted by an unweighted quadratic calibration curve considering ANOVA approach compared to ASTM D5504-12. Despite this, the obtained results show that these two different calibration curve approaches (ASTM and ANOVA) lead to comparable results. Hence, there are no significant statistical differences between these two approaches based on the hypothesis test applied. However, the quadratic calibration curve presents measurement uncertainties of H2S mass fractions much lower than the ASTM approach.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.