comparison of the classical approach using a harmonic signal with the approach of a periodic square wave signal in the first (main) harmonic of the calibration voltage of the electrical power quality analyzers is suggested. The results presented using both approaches for calibrating a Fluke 435 power quality analyzer through a Metrix CX1651 calibrator, provide a quantitative estimate of the degree of equivalence of calibration by both approaches. Evaluation and comparison of the obtained experimental results shows applicability of the approach with a periodic square wave signal of the calibration of power quality analyzers by harmonic voltage in the metrological practice.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.