Abstract
We aimed to compare apparent diffusion coefficient (ADC) values among magnetic resonance imaging (MRI) scanners from different vendors. We used a custom-made phantom solution consisting of distilled water, 0.9% NaCl, 25% NaCl, and shampoo for diffusion-weighted MRI (DW-MRI) examinations. DW-MRI was performed with similar sequence parameters using six different 1.5 Tesla MR scanners (scanners A-F). ADC maps were automatically constructed for all DW-MR images (b factors of 0 and 1000 s/mm(2)). ADC measurements were performed using regions of interest and seven different software programs, including four different postprocessing workstations, two different picture archiving and communication systems, and operator console software for each MR scanner. The ADC values generated by scanners A and F were higher and those of scanner B were lower than those generated by the other scanners (P = 0.002). The intravendor difference in the ADC values averaged from scanners D, E, and F was statistically significant (P < 0.001). The difference between the ADC values obtained by scanners C and E was not statistically different (P = 0.15). ADC values may differ among different MRI systems used for DW-MRI. Thus, the MRI vendor should be considered when using DW-MRI in a clinical setting.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.