In this paper, we present the imaging parameters and compare both mercuric iodide (HgI2) and amorphous selenium (a-Se) films. Using MCNPX code, we designed the film structure and its thickness for the optimized detector in the diagnostic x-ray range. The mercuric iodide film was formed by a wet binder process, while the amorphous selenium film was deposited by physical vapor deposition (PVD). These deposition methods are capable of being scaled up to sizes required in diagnostic imaging applications. The electronic properties are investigated using dark current, x-ray sensitivity and signal to noise ratio (SNR). From our results, the developed HgI2 film as an alternative to a-Se photoconductor, which is in practical use in flat panel x-ray imaging detector, will prove its usefulness in the future design and the optimization for various diagnostic modalities.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.