Abstract
Four closed-vessel microwave digestion methods were compared for the accurate determination of mercury in sediments by flow-injection cold-vapor atomic absorption spectrometry. The results showed that several acid mixtures (HNO3/H2SO4, HNO3/HClO4, HCl/HNO3 and HCl/HNO3/HF) completely digested the soil matrix for the determination of mercury. The method using aqua regia is preferred because it is time saving, less dangerous and suitable for other trace-metal analyses. The merits of pressure-feedback microwave digestion is that it simplifies soil sample digestion, and there is no loss of mercury. The digestion methods were evaluated by determining mercury in NIST SRM 1645 River Sediment, NIES CRM No. 2 Pond Sediment and NRCC BCSS-1 Marine Sediment. Recoveries of 92-108% were achieved. Good recoveries of 94-104% were also obtained for soil and marine-sediment samples spiked with different species of mercury.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.