Abstract

A gold film mercury analyser was used for the determination of mercury in lake sediments. Sediment and soil standards and sediment samples were extracted with a mixture of nitric and hydrochloric acids. Aliquots of the extracts containing between 30 and 50 ng of mercury were used for the determination with the gold film analyser. An absolute deviation of approximately 0.05 µg g–1 and an accuracy of ±0.02 µg g–1 can be considered typical for the investigated method. The results of the determination of mercury in the sediment standard by the gold film analyser were compatible with the recommended value.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.