Abstract

Objective: To evaluate the auditory brainstem response (ABR) amplitudes evoked by tone pip and narrowband chirp (NB CE-Chirp) stimuli when testing post-screening newborns and to determine the difference in estimated hearing level correction values. Design: Tests were performed with tone pips and NB CE-Chirps at 4 kHz or 1 kHz. The response amplitude, response quality (Fmp), and residual noise were compared for both stimuli. Study sample: Thirty babies (42 ears) who passed our ABR discharge criterion at 4 kHz following referral from their newborn hearing screen. Results: Overall, NB CE-Chirp responses were 64% larger than the tone pip responses, closer to those evoked by clicks. Fmp was significantly higher for NB CE-Chirps. Conclusion: It is anticipated that there could be significant reductions in test time for the same signal to noise ratio by using NB CE-Chirps when testing newborns. This effect may vary in practice and is likely to be most beneficial for babies with low amplitude ABR responses. We propose that the ABR nHL threshold to eHL correction for NB CE-Chirps should be approximately 5 dB less than the corrections for tone pips at 4 and 1 kHz.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.