Abstract

The specific absorption rate (SAR) is an indicator of the thermal effects caused by microwave exposure in biological tissue. Conventionally, an electric field probe scanning method is used for SAR evaluation. Although many countries have confirmed the validity of the electric field probe scanning method used in many SAR evaluation devices, a comparison between physically different measurement techniques is an important goal for validating a standard and for evaluating the uncertainty of a measurement system. Hence, the authors developed a thermal SAR measurement method using an optical fiber thermal sensor. In this paper, the results of comparison measurements between the electric field probe scanning system and the proposed thermal SAR measurement system are presented. The results show that the SAR measurements by each system overlap within the range of the expanded standard uncertainty in the frequency range of 2-6 GHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.