Abstract
Main textThe report summarises the results of the EURAMET.T-K8 key comparison (EURAMET Project No. P 717) in the field of humidity. The comparison covers the dew-point temperature range from +30 °C up to + 95 °C dew point temperature. The report gives details of the scheme, procedure and analysis of the comparison. The overall results display a very good agreement of the 16 participating laboratories. The two dew-point hygrometers used as travelling standard were sufficiently stable. The linkage to the corresponding CCT-K8 comparison will be calculated when the final results for both comparisons are available. Possible procedures are described in the text. In the conclusion, some remarks for improvement for future RMO key comparisons are given.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).