Abstract

Specific activity results of (51)Cr, (67)Ga, (99m)Tc and (201)Tl radioactive solutions, measured in different dose calibrators located at nuclear medicine services and in a calibrated HPGe gamma spectrometer, were compared. The HPGe spectrometer was calibrated in a well defined geometry by means of (60)Co, (133)Ba, (152)Eu, (166m)Ho and (241)Am sources, previously standardized in a 4pibeta-gamma coincidence system. Despite the observed differences, the results may be accepted within the 10% uncertainty range, established by Brazilian regulatory standards.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.