Abstract

Considerable inaccuracy and unreliability have been demonstrated to be associated with the widely used radioenzymatic methods for the determination of histamine in biological fluids. Urine appears to inhibit the methylation of histamine by histamine N- methyltransferase such that the radioenzymatic assay underestimates the concentration of histamine present in urine. Directly comparing the radioenzymatic assay with a recently developed reference method using mass spectrometry for the determination of urinary histamine, up to 34-fold differences in the levels of urinary histamine were found with the two methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.